NOSHOK 756 Series Digital Pressure Transmitters User Manual
Page 32

29
7
Diagnostics and Service
The following error messages can appear on devices with displays (see also para.
2.1.3):
8
Disposal
If the fault cannot be corrected, the device should be taken out of operation
and stored to prevent re-installation.
Repairs should only be performed by the manufacturer. All other repairs or
alterations are unauthorized.
s
e
r
u
s
a
e
M
n
o
i
t
c
e
r
r
o
C
r
o
r
r
E
r
o
r
r
E
e
d
o
C
r
o
r
r
E
r
e
r
u
t
c
a
f
u
n
a
m
o
t
e
c
i
v
e
d
n
r
u
t
e
R
r
o
r
r
e
-
M
O
R
0
0
E
E01
Power supply error
Test power supply
E03
EEPROM communications error
Disconnect and reconnect power
supply
E04
Sensor’s temperature range was
exceeded
Return sensor’s temperature to
specified limits
E06
Sensor recognition
Disconnect and reconnect power
supply
E07
General communications error
between the sensor and the con-
trol interface unit
Check the connection between
the sensor and the control inter-
face unit
Please observe local guidlines and regulations when disposing of
devices that are no longer serviceable.
Please turn any recycleable components in to the appropriate local orga-
nizations.
Attention
Attention
Important