9.1 Probe

Care

Under typical operating conditions, the probe should last for several years. Proper care

and maintenance will help you receive the maximum probe life and ensure more accurate

readings.

Since any deposits on the membrane surface act as a barrier to oxygen diffusing through the

membrane, the membrane must be cleaned at regular intervals to assure maximum reliability.

After using the probe, rinse the probe in clean water and wipe it with a soft cloth or paper to

avoid any hardening of deposits. If growth develops on the probe, use a disinfecting chemical

to clean.

NOTE

Although the membrane is strong and not easily damaged, wipe it gently while cleaning it. If

the membrane is damaged or torn, the probe will no longer function.

There are no special probe storage requirements.