10.3.4
Stability setup program P2.2
Use this program to turn on or off the mV stability indicator and
to set level of stability (low, medium or high).
Stability definitions
High stability:
reading is stabilized quickly, but reading is less accurate.
Medium stability:
reading stability is averaged between high and low stability.
Low stability:
reading is stabilized slowly, but guarantees high accuracy
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37).
2. Use
∆/MI key to scroll to program Stability Setup: P2.2. See
Figure 119.
3. Press
ENTER to enter program Stability Setup: P2.2.
4. Press
>/CAL to toggle between mV Stable Criteria: On and
Off. The selection will flash. See Figure 120.
5. Press
ENTER to confirm selection.
•
If you selected On, continue to step 6.
• If you select Off, pressing ENTER returns to the setup
mode.
Press
∆/MI key to continue to the next setup program, or
press
∇/MR to go to the previous program. Press EXIT to
return to the mV measurement mode.
6. If you selected On in step 5, the display now reads Degree of
Stability: Low Medium High. Press >/CAL to toggle between
Low, Medium and High. The selection will flash.
See Figure 121.
7. Press
ENTER to confirm selection and return to the setup
mode. Press
∆/MI to continue to the next setup program, or
press
∇/MR to go to the previous program. Press EXIT to
return to the pH measurement mode.
Degree of Stability
Low Medium High
mV Stable Criteria
On Off
Stability Setup : P2.2
Press ENT. to Select
Figure 119
Figure 120
Figure 121