10.3.4

Stability setup program P2.2


Use this program to turn on or off the mV stability indicator and
to set level of stability (low, medium or high).


Stability definitions

High stability:
reading is stabilized quickly, but reading is less accurate.
Medium stability:
reading stability is averaged between high and low stability.
Low stability:
reading is stabilized slowly, but guarantees high accuracy

1. Enter mV setup mode as described in steps 1-4 in Section

10.3.1 Entering mV setup mode (page 37).


2. Use

/MI key to scroll to program Stability Setup: P2.2. See

Figure 119.


3. Press

ENTER to enter program Stability Setup: P2.2.


4. Press

>/CAL to toggle between mV Stable Criteria: On and

Off. The selection will flash. See Figure 120.


5. Press

ENTER to confirm selection.

If you selected On, continue to step 6.

• If you select Off, pressing ENTER returns to the setup

mode.

Press

/MI key to continue to the next setup program, or

press

/MR to go to the previous program. Press EXIT to

return to the mV measurement mode.


6. If you selected On in step 5, the display now reads Degree of

Stability: Low Medium High. Press >/CAL to toggle between
Low, Medium and High. The selection will flash.
See Figure 121.


7. Press

ENTER to confirm selection and return to the setup

mode. Press

/MI to continue to the next setup program, or

press

/MR to go to the previous program. Press EXIT to

return to the pH measurement mode.

Degree of Stability
Low Medium High

mV Stable Criteria
On Off

Stability Setup : P2.2
Press ENT. to Select

Figure 119

Figure 120

Figure 121