10.4.8

Ion calibration data reset setup program P3.6


Use this program to erase all ion calibration data.


1. Enter ion setup mode as described in steps 1-4 in Section

10.4.1 Entering ion setup mode (page 43).


2. Use

/MI key to scroll to program Ion Cal Reset: P3.6. See

Figure 155.


3. Press

ENTER to enter program Ion Cal Reset: P3.6.


4. Press

>/CAL key to toggle between Ion Reset: Yes and

No. The selection will flash. See Figure 156.


5. Press

ENTER to confirm selection.

• If you selected Yes, the screen flashes Ion being

Reset, Please wait... See Figure 157. The meter then
returns to setup mode.

• If

you

select

No, the screen flashes Reset Aborted and

then returns to the setup mode. See Figure 158.

This is the last ion setup program. Press

/MR to go to the

previous program. Press

/MI to go back to the first ion

setup program. Press EXIT to return to mV measurement
mode.

Mode


Ion being Reset
Please wait. . .

Ion Reset: Yes No
Select & Press ENTER

Ion Cal Reset :P3.6
Press ENT. to Select

Figure 155

Figure 156

Figure 157

Figure 158