Instruction Manual

pH 1100/ 2100

51

10.5.3

Backlight setup mode P4.1


Use this program to turn backlight on or off and set duration time
(0.1 to 10 minutes).

1. Enter meter setup mode as described in steps 1-4 in Section

10.5.1 Entering meter setup mode (page 50).


2. Use

/MI key to scroll to program Backlight Setup: P4.1.

See Figure 167.


3. Press

ENTER to enter program Backlight Setup: P4.1.


4. Press

>/CAL key to toggle between Backlight On and Off.

The selection will flash. See Figure 168.


5. Press

ENTER to confirm selection.

If you selected On, continue to step 6.

• If you select Off, pressing ENTER returns to the setup

mode.


Press

/MI to continue to the next setup program, or press

/MR to go to the previous program. Press EXIT to return to measurement mode.


6. If you selected On in step 5, the display now reads Key BL On

Period, (0.1-10.0 min) =2.00. See Figure 169. Key in the
backlight on period using the numeric keypad. If you make a
mistake, use

∆/MI or ∇MR keys to highlight mistake, then re-

enter .


7. Press

ENTER to confirm selection. The screen flashes Sys

Updating... Please wait... and then returns to the setup mode.
Press

/MI key to continue to the next setup program, or

press

/MR to go to the previous program. Press EXIT to

return to measurement mode.

BackLight Setup:P4.1

Press ENT. to Select

BackLight On Off

Select & Press ENTER

Key BL ON Period
( 0 . 1-10 . 0 min) = 2.00

Figure 167

Figure 168

Figure 169