10.4.5

Stability setup program P3.3


Use this program to turn on or off the ion stability indicator and
to set level of stability (low, medium or high).

Stability definitions

High stability:
Reading is stabilized quickly, but reading is less accurate.
Medium stability:
Reading stability is averaged between high and low stability.
Low stability:
Reading is stabilized slowly, but guarantees high accuracy.

1. Enter ion setup mode as described in steps 1-4 in Section

10.4.1 Entering ion setup mode (page 43).


2. Use

/MI key to scroll to program Stability Setup: P3.3 See

Figure 141.


3. Press

ENTER to enter program Stability Setup: P3.3


4. Press

>/CALkey to toggle between Ion Stable Criteria: On

and Off. The selection will flash. See Figure 142.


5. Press

ENTER to confirm selection.

If you selected On, continue to step 6.

• If you select Off, pressing ENTER returns to the setup

mode.


Press

/MI key to continue to the next setup program, or

Press

/MR to go to the previous program. Press EXIT to

return to the ion measurement mode.


6. If you selected On in step 5, the display now reads Degree of

Stability: Low Medium High. . Press >/CAL to toggle between
Low, Medium and High. The selection will flash. See Figure
143
.


7. Press

ENTER to confirm selection and return to the setup

mode. Press

/MI key to continue to the next setup program,

or press

/MR to go to the previous program. Press EXIT to

return to the ion measurement mode.

Ion Stable Criteria
On Off

Degree of Stability
Low Medium High

Stability Setup : P3.3
Press ENT. to Select

Figure 141

Figure 142

Figure 143