2 ph calibration – Thermo Fisher Scientific Alpha pH 500 User Manual

Page 14

Advertising
background image

10

5.2 pH Calibration

A complete 2-point CAL is required for this Transmitter. If Calibration is aborted
after 1st calibration point, transmitter reverts to previous Calibration data.

1. Rinse probe thoroughly with de-ionized water or a

rinse solution. Blot excess liquid.

2. Dip probe into calibration buffer. End of probe

must be completely immersed into buffer. Stir

probe gently to create a homogeneous sample.

3. From pH MEAS mode, press CAL to enter pH

calibration mode. CAL indicator will be shown.
Primary display will show measured reading while
secondary display will indicate the pH standard

buffer solution.

4. Press ▲ or ▼ keys to select either pH 7.00 (USA)

or pH 6.86 (NIST) standard buffers. This is the first point, Offset

calibration.

5. Wait for measured pH value to stabilize. Press

ENTER to confirm calibration. Transmitter is now

calibrated to current buffer.

6. Rinse probe with de-ionized water or rinse

solution, and place in next pH buffer (pH 4.01,

pH 9.18 or pH 10.01).

7. Repeat steps 5 and 6.
8. Upon successful Calibration, transmitter displays

Slope (primary display, in mV) and Offset

(secondary display, in pH) of electrode.

9. Press ENTER to return to MEAS mode.

Transmitter has been calibrated. Rinse electrode

and proceed with measurements.

NOTES:
To exit from pH calibration mode without confirming calibration, press ▲ and ▼
keys together.
If selected buffer value is not within

1.0 pH from measured pH value: electrode

and buffer icon blink and ERR annunciator appears in lower left corner of display.

CAL

HOLD

pH

READY

pH

CAL

HOLD

pH

READY

pH

CAL

HOLD

pH

READY

pH

CAL

HOLD

pH

mV