5.3 Relative mV Calibration

1. From the ORP MEAS mode, press CAL key.

CAL indicator appears at top of LCD. Primary
display shows relative mV reading and

secondary display shows absolute mV value.

2. NOTE: If you have never calibrated relative mV

or if the transmitter has been reset, the value
shown in the primary display is the same as the

absolute mV value.

3. Press ▲ or ▼ keys to enter the relative mV

value that matches your desired reading (only

the readings in the primary display changes).

4. Press ENTER key to confirm. Transmitter will

now display the offset in the readings.

5. Press ENTER key to return to MEAS mode.

5.4 Temperature Calibration

Temperature calibration is available only in the pH measurement mode and when

ATC probe is available.
Calibrate temperature probe only if temperature value displayed on transmitter is
different from that of a calibrated thermometer. Please refer to section 6.3.1 for

further information.