SETUP

HOLD

6.2 Output Range Setting

Output of pH 500 transmitter is via the 2-wire power supply
loop. Both the 4 mA and 20 mA output can be assigned to

specific pH or ORP values, for a more refined output.

1. Press ENT key and use ▲ or ▼ keys to scroll till LCD

displays „

r n g ‟. Press ENT again. Upper display

shows pH value, „0.00 pH‟ (or – 1000 mV) while lower

display shows „

r 4.0‟. This is the set up page for the

pH or ORP output range setting for the 4mA output

current.

2. Assign pH or ORP value to 4 mA output current. Use

▲ or ▼ keys to change pH or ORP value. Press ENT

to accept value.

3. Upper display shows pH value,‟

14.00 pH‟ (or 1000 mV) while lower display

shows „

r 20.0‟. This is the set up page for the pH or ORP output range

setting for the 20mA output current.

4. Assign pH or ORP value to 20 mA output current. Use ▲ or ▼ keys to

change pH or ORP value. Press ENT to accept value.

5. Press both ▲ or ▼ keys together to return to „ MEAS „ mode.