18

6.6 Configuration

6.6.1

Selecting pH or ORP Mode of Operation

The alpha pH 500 transmitter is not only a pH transmitter, but
also an ORP transmitter. When configured to measure ORP,
electrode must be changed and calibrated (see Section 5.3 for
calibration procedure). ORP values are measured and

displayed in mV only.
1. Press ENT key and use ▲ or ▼ keys to scroll till LCD

displays „

COF

(Configuration). Press ENT again.

2. LCD displays „

PH

. Transmitter is set to operate as a pH

Transmitter. Use ▲ or ▼ keys to change from „

PH

to

ORP

, if this Transmitter is to operate as an ORP

Transmitter.

3. Press the ENT key to accept selection.
4. Press  and  keys together, to return to MEAS mode.

5. If the readings on the Transmitter is fluctuating wildly, please proceed to

Section 6.6.2, for a possible solution.

6.6.2

Selecting Asymmetrical or Symmetrical Mode of Operation

In some applications, there may be some electrical interference in the sample,
where the electrode is installed. The sensitive electrode picks up the signal and the

effect is displayed on the LCD.
If the electrical interference is from an AC source, the readings on the LCD
fluctuates wildly. If the source is DC, then readings will be stable, but at a wrong

value.
In such cases, switch the Transmitter to the Symmetrical mode as below:

1. Press ENT key and use ▲ or ▼ keys to scroll till LCD

displays „

COF

(Configuration). Press ENT again.

2. LCD displays either „

PH

or „

ORP

depending on

requirement. Press ENT key.

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD