6.4 HOLD Current Setting

When Transmitter is in CAL or SETUP modes, it automatically
goes into a „HOLD‟ mode.
To indicate Transmitter is in „HOLD‟ mode, output current can be

set to 22 mA output by activating the „HLD On‟.

1. Press ENT key and use ▲ or ▼ keys to scroll till LCD

displays „SEt‟ in the upper display; and „HLd‟ in the

lower display. Press ENT again.

2. Upper display now shows „HLd‟. Lower display will

show either „OFF‟ or „On‟. Use ▲ or ▼ keys to toggle

between „On‟ of „OFF‟. Press ENT to accept selection.

3. Press  and  keys together, to return to MEAS

mode.

NOTE: If „HLd‟ is set to „OFF‟, then current output will be
equivalent to last measured value.

6.5 Out-of-Range Current Setting

If pH or ORP readings exceed the scale, LCD displays „Or‟ (over-
range), or „Ur‟ (under-range). Transmitter current output can be

set to 3.8 mA to indicate Out-of Range situation.

1. Press ENT key and use ▲ or ▼ keys to scroll till LCD

displays „SEt‟ in the upper display; and „org‟ in the lower

display. Press ENT again.

2. Upper display now shows „org‟. Lower display will show

either „OFF‟ or „On‟. Use ▲ or ▼ keys to toggle

between „On‟ of „OFF‟. Press ENT to accept selection.

3. Press  and  keys together, to return to MEAS mode.

NOTE: If „org‟ is set to „OFF‟, under-range will be set at 4 mA,
while over-range will be set at 20 mA.