2 Instrument Description

CF card full

The signal is activated when the memory space of a CF card
that has been inserted (external memory) has fallen below the
selected threshold value.

v Chapter 2.10 “Reading out Data”

Mem.al.
interface

The signal is activated when the device-internal memory
space available for data readout via interface has fallen
below the selected threshold value.

v Chapter 2.10 “Reading out Data”

(Mem.al. = memory alarm)

Login

The signal is activated as soon as a user is logged in to
instrument.

v Chapter 8.2 “Log-in and Log-out”

Error

The signal “Error” (also referred to as system error) is
activated when:

- the CF card is faulty or full,

- the device battery is empty,

- the internal memory is faulty or

- a p.c.b in the device has not been calibrated.

Reserve 1

Currently not used.

Fieldbus
error

The signal is activated when an error occurs during
communication via PROFIBUS.

Reserve 2

Currently not used.

CF card
inserted

The signal is activated when a CF card is inserted in the
instrument.

CF card
removed

The signal is activated when a CF card is removed from the
instrument.

Signal

Description