Troubleshooting, Troubleshooting 12 – Badger Meter Test Equipment User Manual

Page 12

Advertising
background image

TROUBLESHOOTING

Problem

Possible Causes

Remedies

Screen is blank

Battery is dead

Replace battery

Will not count in totalizer mode

1 Improperly programmed
2 Broken or defective wiring
3 Improperly connected
4 Transmitter defective

1 Check programming
2 Check wiring
3 Check connections
4 Repair or replace transmitter

Will not indicate rate of flow

1 Improperly programmed
2 Improperly connected
3 Transmitter defective

1 Check programming
2 Check connections
3 Repair or replace transmitter

Cannot program

Program-enable jumper is not installed or

is installed improperly

Install jumper from Terminal 1 to

Terminal 5

Cannot reset from front panel

Reset enable is not programmed

Reprogram mode to R

(Program Mode 6)

Erroneous readings

1 Improperly programmed
2 Defective transmitter

1 Check programming
2 Repair or replace transmitter

Table 4: Troubleshooting

For further assistance, call our Technical Support Staff at 1-800-876-3837

Model ER-10, ER-10/R & ER-10H Industrial Register

Page 12

September 2013