CALIBRATION

OTEE:

N

If your Intelligent Frequency Converter was purchased with a B220-954 turbine meter, the two components ship

from the factory calibrated as a set If the Intelligent Frequency Converter is a replacement, the turbine’s K factor has

changed, or the converter is being used with some other pulse-generating device, programming will be necessary

Requirements:
• B220-954 Intelligent Converter Programming Kit (B220-954)
• IBM Compatible PC running Windows 95 and newer
• DC Power Supply
After installing the programming software and attaching the programming cable to the Intelligent Frequency Converter

through an available Com Port on your PC, you are now ready to power up your Intelligent Frequency Converter and continue

with the calibration procedure Using a DC power supply, apply 10…26V DC to the Intelligent Frequency Converter Run the

programming utility software shown in

Figure 6

IFC Programming Utility

Setup

Frequency

Device:

Rate Units:

Rate Interval:

K Factor Units:

K Factor:

Damping:

Flow at 4mA:

Flow at 20mA

Linear Points:

Download Setup

Read Setup

Linear Points

Status: Com1

9/23/2008

2:33 PM

0 0 0 0 Hz

Monitor

On
Off

File Options Tools

4-20mA

0-5V

Pulses/Gallon

350.400

0

50.000

0.000

0

Gallons

Minute

IFC Programming Utility

Setup

Frequency

Device:

Rate Units:

Rate Interval:

K Factor Units:

K Factor:

Damping:

Flow at 4mA

Flow at 20mA

Linear Points:

Download Setup

Read Setup

Linear Points

Status: Com1

9/23/2008

2:33 PM

0 0 0 0 Hz

Monitor

On
Off

File Options Tools

4-20mA

0-5V

Pulses/Gallon

350.400

0

50.000

0.000

0

Gallons

Minute

Figure 6: IFC programming software

Pressing Read Setup will read the current configuration of your sensor After entering your desired values for Device,

K factor, rate and flow span, press Download Setup to save your settings Please consult the factory for any further

calibration requirements

OTEE:

N

If communication fails, please check cabling and/or ComPort address and try again

OTEE:

N

For complete programming instructions refer to the IFC Programming Kit manual