MAINTENANCE

General

This section contains information for servicing and maintaining of the meter The information consists of preventive

maintenance, calibration and adjustment and general servicing procedures Instructions for ordering replaceable parts

and assemblies are provided in an Illustrated Repair Parts Bulletin that is included with this manual Any service and repair

procedures applicable to a meter-mounted or free-standing accessory are provided in the bulletin or instruction manual

covering the device

Preventive Maintenance

The purpose of preventive maintenance is to ensure efficient operation and long life by detecting and correcting any defects

before damage or meter failure occurs Preventive maintenance consists of periodic inspection and cleaning procedures

The procedures should be performed at regular intervals and any defects discovered should be corrected before attempting

further operation

1 Visually inspect the meter and applicable accessory for loose connections, missing hardware, cracked or broken dial

glass, damaged wiring or any other signs of wear or deterioration Repair or replace parts as necessary

2 Clean all dust, dirt or other foreign material from the exterior of the meter and accessory
3 Fluid metering applications involving food ingredients require that Only manual cleaning (remove meter housing

cover) can be used for sanitary OP meters

4 Fluid metering applications involving chemical solutions can be cleaned manually or mechanically by a meter bypass

and flush system

If the facility piping arrangement includes a meter bypass and flushing system (see

"Figure 2"

), shut off fluid flow to the meter

and flush interior of meter housing with clean water to remove any buildup of deposits or corrosion If a bypass and flushing

system is not included in the metering application, remove the meter housing cover and inspect the measuring chamber,

piston and associated parts for any buildup of deposits or corrosion Clean and flush parts with water or a suitable solvent that

will not adversely affect the plastic components

Calibration Check and Adjustments

All meters are accuracy tested with water at the factory using calibrated test equipment However, because the meter can

be used to measure a wide variety of liquids that may vary from water in viscosity, it is sometimes necessary to recalibrate a

meter under actual operating conditions with the liquid being metered
The following instructions are provided to assist in performing an on-site calibration check and adjustment

Accuracy Test

1 Place a test tank of calibrated volume at the output of the meter
2 Operate meter until test tank is filled to the appropriate calibrated level Since meter accuracy varies somewhat with

flow rate, make test run at the same flow rate used in actual operation

3 Record quantity indicated on applicable meter accessory device
4 Repeat run three times and average recorded indications
5 Perform the following calculation to determine the percent of accuracy of the meter

Example 1:

In this example, the meter accessory is slow and must be speeded

up by a calibration correction

Example 2:

In this example, the meter accessory is fast and must be slowed

down by a calibration correction

Industrial Meter, Oscillating Piston, Sizes 1/2" through 2" Model OP

Page 10