Low Battery Indication

The Battery icon ( ) uses three bars to display the battery level. When the icon displays all three bars, the batteries are full. The XP2i will continue to operate
accurately while the icon is visible. When the batteries are exhausted, the letters

batt

will appear across the display. After

batt

appears, no pressure measure-

ments will be possible until the batteries are replaced.

Battery replacement

The XP2i uses 3 AA batteries. Loosen the four rear panel screws to gain access to the battery compartment. These four panel screws are captive inside the
panel and are not removable. Pull the panel up to expose the battery compartment. After replacing the batteries, the XP2i will start operating immediately
(without having to press the

(

on/off

)

button). This indicates that a complete reset has occurred, and is normal.

!

WaRninG: do not remove or change the batteries in hazardous locations.

!

WaRninG: t4 or t3 temperature class is based on selection of approved battery. see the table below.

!

Caution: do not mix battery types or manufacturers.

X

the XP2i is Intrinsically safe only if powered by one of the following battery types:

approved Battery type

ta=

Marking

Rayovac Max Plus 815

-20 to 50° C

Ex ia IIC T4 Ga

Energizer E91

*

Energizer EN91

*

-20 to 50° C

Ex ia IIC T3 Ga

Duracell MN1500

*

Energizer is manufactured by Energizer Holdings, Inc., and the Eveready Battery Company, Inc.

Many other battery types and models have been tested but failed to meet the requirements for Intrinsic Safety—do not assume other models are equivalent.
The XP2i can be operated from an external power supply (AC adapter kit PN: 2984).

!

WaRninG: do not use ac adapter in hazardous locations.

X

the XP2i is csa certified only if powered by one of the following battery types:

approved Battery type

ta=

Marking

Rayovac Max Plus 815

-20 to 45° C

Class I, Division 1, A, B, C, D T4

Energizer E91

-20 to 50° C

Class I, Division 1, A, B, C, D T3C

Energizer EN91

Rayovac Max Plus 815

!

WaRninG: use of duracell Mn1500 batteries is not approved for csa certification.