Specifications

26

XP2i Operation Manual

Pressure conversions

1 psi =

27.6806 inches of water column (water at 4° C [39.2° F])

27.7070 inches of water column (water at 15.6° C [60° F])

27.7292 inches of water column (water at 20° C [68° F])

2.03602 inches of mercury (mercury at 0° C [32° F])

51.7149 millimeters of mercury (mercury at 0° C [32° F])

703.087 millimeters of water column (water at 4° C [39.2° F])

0.070307 kilograms per square centimeter

68.948 millibar

6.8948 kilopascals

0.068948 bar

0.006895 megapascals

connections

Communication . . . . . . . . . . . . . . . . . DB-9, RS-232 (environmentally sealed).

!

WaRninG: do not use rs-232 in hazardous locations.

Pressure Connection . . . . . . . . . . . . Crystal CPF Female (1/4" medium pressure tube system).

Compatible with HiP LM4 and LF4 Series, Autoclave Engr SF250CX Male and Female Series.

CPF o-ring size and material: AS568A-012, Viton 90 durometer (P/N 3981).

!

Caution: to achieve cPf maximum allowable working pressures no o-ring substitutions are allowed. see

crystal engineering’s cPf brochure

and

ces-003 cPf safety Instructions

for further detail.

display

Screen . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5 digits.

Display Rate . . . . . . . . . . . . . . . . . . . . . 4 readings/second (standard).

8 readings/second (PSV mode).

Numerical Display Height . . . . . . . . 16.9 mm (0.67”) single line display. LCD readable in sunlight

with bright backlight.

sensor

Wetted Materials . . . . . . . . . . . . . . . . (Wrench Tight) 316 stainless steel

(Finger Tight) 316 stainless steel and Viton (internal o-ring)

Diaphragm Seal Fluid . . . . . . . . . . . . Dow Corning 200