Specifications – Crystal DataLoggerXP Digital Pressure Gauge Software User Manual

Page 27

Advertising
background image

Specifications

25

XP2i Operation Manual

Specifications

accuracy

Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year.

Gauges must be exercised whenever exposed to significant changes in environmental conditions to achieve these specifications, and (if not an absolute
model), rezeroed. To exercise a gauge, cycle the gauge between zero (ambient barometric pressure) and the pressure of interest. A properly exercised gauge
will return to a perfect zero reading (or return to the same ambient barometric reading).

Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertification period.

20 to 100% of Full Scale . . . . . . . . . . ±(0.1% of Reading).

0 to 20% of Full Scale . . . . . . . . . . . . ±(0.02% of Full Scale).

Vacuum, for 300 psi, 20 bar, 2000 kPa, and lower pressure gauges:

±(0.25% of Full Scale), where F.S. = -14.5 psig, -1.0 bar, -100.0 kPa.

!

CAUTION: Not recommended for continuous use below -14.5 psig or 0.2 psiA. Refer to the

XP2i-dP

data sheet for gauges that are intended for

continuous high vacuum use.

temperature

Operating . . . . . . . . . . . . . . . . . . . . . . . -10 to 50° C (14 to 122° F).

Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification.

Storage. . . . . . . . . . . . . . . . . . . . . . . . . . -40 to 75° C (-40 to 167° F).

Battery should be removed if stored for more than one month.

Media compatibility

Liquids and gases compatible with 316 Stainless Steel.

Advertising
This manual is related to the following products: