Dwyer SBLTX User Manual

Mercoid division, Series sbltx submersible level transducer

Advertising
background image

Series SBLTX Submersible Level Transducer

Specifications - Installation and Operating Instructions

Bulletin L-37

The SBLTX Submersible Level Transducer is manufac-
tured for years of trouble free service. The transducer con-
sists of a piezoresistive sensing element, encased in a 316
SS housing. Bullet nose design protects diaphragm from
damage. Comes equipped with a 270-pound tensile strength,
shielded, vented cable. Ventilation tube in the cable automat-
ically compensates for changes in atmospheric pressure
above the tank.

Intrinsic Safety Approval Classification
The SBLTX is UL listed for use in Hazardous (Classified) Loca-
tions. The protection method is by Intrinsic Safety, “ia”. It was in-
vestigated by UL under UL Standard 913 Sixth Edition and CSA
Standard No. 157-92.

For use in Hazardous (Classified) Locations:
Class I Div. 1 Groups A,B,C,D
Class II Div. 1 Groups E,F,G
Class III Div. 1
Temperature Code: T4 @ 80°C ambient
Install in accordance with control drawing 01-700797-00.

WARNING: To prevent ignition of flammable
or combustible atmospheres, disconnect
power before servicing.
Use with approved safety barriers using entity
evaluation.

SPECIFICATIONS
Service:
Compatible liquids.
Wetted Materials: 316 SS, 316L SS,

Buna-N; Cable:

Polyurethane or ETFE; Bullet Nose: PVC.
Accuracy: ±0.25% of full scale.
Temperature Limit: 0 to 176°F (-18 to 80°C).
Compensated Temperature Range: 0 to 176°F (-18 to 80°C).
Thermal Effect: Less than ±0.02% full scale/ °F.
Pressure Limit: 2X full scale.
Power Requirement: 10 to 28 VDC.
Output Signal: 4 to 20 mA DC, 2-wire.
Response Time: 50 ms.
Max. Loop Resistance: 900 ohms.
Electrical Connections: Wire pigtail.
Mounting Orientation: Suspended in tank below level being
measured.
Weight: 2.2 lb (1.0 kg).
Agency Approvals: CE, UL Intrinsically Safe to UL Standard
913. (See Intrinsic Safety Approval Classification.)
The following standards were used for CE approval:
IEC 61000-4-2: 2001
IEC 61000-4-3: 2006
IEC 61000-4-4: 2004
IEC 61000-4-5: 2005
IEC 61000-4-6: 2006
IEC 61000-4-8: 2001
CENELEC EN 55011: 2003
CENELEC EN 61326: 2003
89/336/EEC EMC Directive

MERCOID DIVISION

DWYER INSTRUMENTS, INC.
P.O. BOX 258 • MICHIGAN CITY, IN 46361 U.S.A.

Phone: 219/879-8000

www.dwyer-inst.com

Fax: 219/872-9057

e-mail: [email protected]

BREATHER TUBE

SUBMERSIBLE
CABLE

1/2 NPT

6-15/16

[176.21]

Ш1

[Ш25.40]

Vmax

28VDC

Imax

93mA

Ci

0.051_F

Li

240_H

Entity Parameters

Pi

0.651W

Project7:Layout 7 12/7/09 12:01 PM Page 1

Advertising