Wiring, E-m-hf5-v1_25 – ROTRONIC HF5 User Manual

Page 16

Advertising
background image

E-M-HF5-V1_25

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroFlex HF5 Humidity Temperature
Transmitters: User Guide

Instruction Manual

Document Type

Page

16 of 30

Document title

© 2009-2013; Rotronic AG E-M-HF5-V1_25

6.4 Wiring

6.4.1

HF52: 2-wire, loop powered transmitter


Electrical diagram

The maximum permissible cable length connecting
the HF52 to other devices is determined by the total
resistance resulting from the addition of the cable
resistance and that of the devices connected in series
with the unit. This resistance should not exceed 500
ohms.






Terminal block diagram




Note: connect the + of the power supply to only one of the V+ terminals. The two terminals marked V+ are
internally connected.

Measuring humidity or temperature only

Unless configured to measure either humidity only or temperature only, proper operation of the HF52 requires
both current loops to be closed. The HF52 can be directly ordered from the factory to measure either humidity
or temperature only. When necessary, any unused output of the HF52 can be disabled with the ROTRONIC
HW4 software. When the HF52 is configured with one of the two outputs disabled, close only the loop that is
being used.

Terminals

Description

K2-2: T-OUT

Temperature output (+)
OUT-2

K2-1: V+

Power supply: 10…28 VDC (+)

Terminals

Description

K1-2: H-OUT

Relative humidity or dew point (+)
OUT-1

K1-1: V+

Power supply: 10…28 VDC (+)

Advertising