E-m-hf5-v1_25 – ROTRONIC HF5 User Manual

Page 23

Advertising
background image

E-M-HF5-V1_25

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroFlex HF5 Humidity Temperature
Transmitters: User Guide

Instruction Manual

Document Type

Page

23 of 30

Document title

© 2009-2013; Rotronic AG E-M-HF5-V1_25

HF53 and HF55 transmitters:

Main Menu

Menu Items

Selections / Information

Notes

Device Settings

Units

Metric / English

Back Light

Key Press / On / Off

Display backlight mode

Contrast

LC display contrast adjustment

Trend

On / Off

Trend indication on the display

Device Information

Version

Firmware version

Serial Nbr

Serial number

Address

RS-485 address

Type

Device type

Name

Device name

User defined

Probe Information

Version

Firmware version

Serial Nbr

Serial number

Address

RS-485 address

Name

Device name

User defined

SensorTest

Humidity sensor status

Off / Good / SQ-Tuned / Bad

Record

On / Off

Data recording by the probe
(max. 2000 values)

Humidity Adjust

RefValue

Humidity reference value

± 0.1 %RH steps

Acquired

Number of cal. points in probe
memory

<Acquire>

Save cal. point to probe memory

<Delete>

Erases all calibration points

<Adjust>

Effect depends on number of
calibration points

Temperature Adjust

RefValue

Temperature reference value

± 0.1 ˚C steps

<Adjust>

1-point adjustment only (offset)


o

Record: both the recording mode (start / stop and the log interval cannot be changed from the menu and
are as configured with the ROTRONIC HW4 software

o

SensorTest: Off means that the humidity sensor has not been tested due to the configuration settings of
the test. For a description of the automatic humidity sensor test and drift compensation (SQ-tuning) see
documents E-T-AC3000-DF-V1 and E-M-HW4v3-F2-005