7.5 Temperature alarm output

7.5.1 Temperature alarm output: Compare the measurements with the setting value of the

temperature alarm output and turn on and off the temperature alarm output.


• Operation table of temperature alarm outputs

1. Deviation top/bottom limit

5. Absolute value top/bottom limit

E*C

SV

E*C

E*L

E*H

0℃( v

E*C

E*C

E*L

E*H

2. Deviation top limit

6. Absolute value top limit

SV

E*C

E*H

0℃( v

E*C

E*H

3. Deviation bottom/limit

7. Absolute value bottom limit

E*C

SV

E*L

0℃( v

E*C

E*L

4. Deviation top/bottom limit range

8. Absolute value top/bottom limit range

E*C

SV

E*L

E*H

E*C

E*C

E*C

0℃( v

E*L

E*H

Operation range of temperature alarm output: E*L, bottom limit setting of temperature alarm output; E*H, top
limit setting of temperature alarm output; E*C, temperature alarm sensitivity


7.5.2 Additional functions

• Standby sequence
This function inhibits temperature alarm output even if the conditions are met for temperature

alarm output when this product is turned on. Temperature alarm output is generated only when
this product deviates from the conditions for temperature alarm output and satisfies those
conditions again. This function is used for bottom temperature alarm output when this product
is turned on. This function is enabled when this product is turned on in the shaded portion in
the operation table of temperature alarm outputs.

[How to cancel this function]
This function is canceled when this product is activated or when either of the following setting is

changed: target, temperature alarm, PV correction, and temperature alarm setting.

• Temperature alarm output holding
When a temperature alarm output is generated, this function retains that state. Even if this

product deviates from the conditions for temperature alarm output, the temperature alarm output
will remain on.

[How to cancel this function]
To cancel this function, turn this product off and back on or disable the additional function for

temperature alarm output setting.