KANE900 Plus Operators Manual

Page 30

AMBIENT :

Temperature measured by the internal sensor, used in the NET

(Ta)

temperature calculation if an INLET probe is not fitted.


CO/CO2 R :

The CO/CO2 ratio, is the ratio of measured CO divided by

calculated CO2.


It gives an indication of the following :-


How good a gas sample the instrument is reading.

How clean the boiler is running.


For example : A new or clean domestic boiler will display a ratio of
less than 0.004, a unit in need of cleaning 0.004-0.008 and a unit in
need of major overhaul will show greater than 0.008.


This only shows a reading when a combustion test is being carried
out. ‘- -’ is displayed while in clean ambient air.


P INDEX :

The CO/CO2 ratio expressed as a percentage %, called the ‘Poison

(PI)

Index’ i.e. P INDEX % = 100 x CO/CO2. ‘- -’ is displayed while in

clean ambient air.


XAIR % :

Excess air calculated from the measured oxygen and type of fuel

(

λ)

used. During a combustion test ‘O2 > 20%’ will be displayed while

in clean ambient air.



Prs :

Flue draught pressure reading. Displayed when pressure sensor
fitted. See UNITS menu 5.2.3. for scales.


NO:

Nitric Oxide reading in ppm or mg/m3. Displayed when Nitric
Oxide sensor fitted. Also displayed as NO (n) when referenced to
oxygen. The display will read ‘O2 > 20%’ if referenced values
selected and instrument is in clean ambient air.


NOx :

Calculated total Nitric oxides displayed in ppm or mg/m3. Where
NOx = NO + P%NO, note P can be set from 0-9%, default = 5%.
See SELECT menu 5.2.2. Also displayed as NOx (n) referenced to
oxygen. The display will read ‘O2 > 20%’ if referenced values are
selected and instrument is sampling clean ambient air.


SO2 :

Sulphur Dioxide reading in ppm or mg/m3. Displayed when Sulphur
Dioxide sensor fitted. Also displayed as SO2 (n) referenced to
oxygen. The display will read ‘O2 > 20%’ if referenced values
selected and instrument is in clean ambient air.


O2 ref % :

Toxic gas measurements can be referenced to defined oxygen levels.

(O2r)

See SELECT menu 5.2.2 for details.