KANE900 Plus Operators Manual

Page 5

2. SAFETY

WARNING


This analyser extracts combustion gases that may be toxic in relatively low
concentrations. These gases are exhausted from the side of the instrument. This
instrument must only be used in well ventilated locations. It must only be used by
trained and competent persons after due consideration of all the potential hazards.

Protection Against Electric Shock (in accordance with EN 61010-1 :
1993)


This instrument is designated as Class III equipment and should only be connected to
SELV circuits. The battery charger is designated as:

Class II equipment
Installation category II
Pollution degree 2
Indoor use only
Altitude to 2000m
Ambient temperature 0

°C-40°C

Maximum relative humidity 80% for temperatures up to 31

°C decreasing linearly to

50%RH at 40

°C

Mains supply fluctuations not to exceed 10% of the nominal voltage.

3.

FIRST TIME USE


Charge the battery for 12 hours. Following this, an overnight charge should be
sufficient for an average 8 hour day. See Main Parameter displays for Battery
Indicator.

The KANE900 Plus has a rechargeable lead acid battery which uses a different
charger than other Kane analysers. Ensure the correct charger is used or damage
may occur to the instrument.

Check that you have all the items you have ordered.
Take time to read this manual fully. Be aware that the analyser configuration that
you have purchased may not support all the features detailed in this manual.

When using the analyser for the first time you will need to choose from:-
Language

selection

Calibration

countdown

time

CO

gas

alarm

NOx percentage for calculation

Time

and

Date

Printed header name and telephone number


The SET UP MENU (Section 5.2.5) gives details of how to change the above settings.