Co² calibration calibration mode – UEi Test Instruments AQM4 User Manual

Page 16

Advertising
background image

The AQM4 allows you to calibrate CO² and

RH, for CO calibration please send unit to

UEi. To enter calibration mode, place unit

at suitable calibration site specified in the

following sections.

Press + , then also

simultaneously for 3 seconds to turn on the

meter and enter calibration mode (Fig. 33).

Press or to select parameter to

calibrate. To cancel or exit calibration at any

time, turn meter off.

NOTE:

ENSURE THE BATTERIES ARE AT FULL VOLTAGE

BEFORE THE CALIBRATION TO PREVENT

INTERRUPTION OR FAILED CALIBRATION.

The manual calibration is suggested to

be done in sunny outdoor air that is well

ventilated. To begin, enter calibration mode.

Press or to select parameters

for calibration and press enter. The meter

is calibrated at standard 400ppm CO²

concentration at the factory. It is suggested

to manually calibrate regularly to maintain

good accuracy.

NOTE:

WHEN THE ACCURACY BECOMES A CONCERN,

RETURN TO UEi FOR STANDARD CALIBRATION.

CAUTION:

Do not calibrate the meter in the air with

unknown CO² concentration. It could be

calibrated as 400ppm by default which

leads to inaccurate measurements.

Wait 10 minutes until the blinking stops and

the calibration is completed. To cancel the

calibration, turn off the meter at any time.

15

CO² CALIBRATION

CALIBRATION MODE