RH CALIBRATION

To begin, enter calibration mode. Press

or to select parameters for calibration

and press enter.

The meters default for calibration of

humidity is 33% and 75% salt solution. The

ambient condition is recommended at 70 ~

75°F (25°C) and stable humidity. To cancel

calibration, turn off the meter.

CAUTION:

Do not calibrate the humidity without

default calibration salts. Contact UEi for

calibration salt or for service

33% Calibration

Plug the sensor probe into 33% salt bottle.

Hold down to enter 33% calibration.

"CAL" and calibrating value (32.8% if at

77°F) are blinking on the LCD.

75% Calibration

Plug sensor probe into 75% salt bottle and

enter calibration mode. In calibration mode,

press to select 75% calibration (Fig. 35),

press to start. "CAL" calibrating value

(75.3% if at 77°F) will blink on the LCD.

Meter will begin calibration and finish in

60 minutes when "CAL" and humidity stop

blinking. To cancel the calibration, turn off

the meter at any time.