UEi Test Instruments AQM4 User Manual

Page 18

Advertising
background image

When the accuracy becomes a concern,

return to UEi for standard calibration. It

is necessary to have standard CO gas for

calibration.

CAUTION:

Do not calibrate the meter in the air with

unknown CO concentration. It could be

calibrated as 400ppm by default which

leads to inaccurate measurements.

Will Not Power On

- Press for more than 3 seconds.

- Check that the adaptor is connected

and on.

- Check that battery level is sufficient

and properly installed.

Readings Not Changing

- Check whether data hold function is

activated (HOLD icon is at the left top).

Slow Response

- Check the air flow channels on

the rear. Verify no blockage exists.

Error Messages

EO1/E33: CO² sensor is out of order. Try new

battery first.

E02:

The value is under range.

E03:

The value is over range

E04:

The original data error results in

this error (DP, WB)

CO CALIBRATION

TROUBLESHOOTING

17