Features and Functions

2

U2761A User’s Guide

85

Triggering

Applies to sweep only. You can issue triggers for sweeps
using Internal (immediate) triggering, External triggering, or
Manual (software) triggering.

• Internal triggering is enabled when you turn on the

U2761A. In this mode, the U2761A outputs continuously
when the sweep mode is selected.

• External triggering uses the device Trig In connector to

control the sweep. The U2761A initiates one sweep each
time Trig In receives a TTL pulse. You can select whether
the U2761A is triggered on the rising or falling edge of
the External trigger signal.

• Manual triggering initiates one sweep each time you click

Trigger

on the soft front panel.

Trigger Source Choices

You must specify the source from which the U2761A accepts
a trigger.

• The sweep trigger source consists of Internal, External, or

Manual. The default trigger source is Internal.

• The U2761A accepts a Manual trigger, a hardware trigger

from the device Trig In connector, or continuously output
sweeps using an Internal trigger. At power

-on, Internal

trigger is selected.

• The trigger source setting is stored in volatile memory

where the source is set to Internal (immediate) trigger
upon power

-off or after a remote interface reset.

Soft Front Panel Operation

On the Trigger Setup panel as shown in

Figure 2- 28

, select

the desired trigger source from the drop down list

at the Source panel.