Output Termination

This configuration applies to output amplitude and offset
voltage only. The U2761A has a fixed series output
impedance of 50

Ω to the device output connector. If the

actual load impedance is different from the specified value,
the amplitude and offset levels will be incorrect.

• The range of the output termination is 1

Ω to 10 kΩ, or

Infinite. The default value is 50

Ω

.

• The output termination setting is stored in volatile

memory and upon power

-off or after a remote interface

reset, the setting will return to a default value.

• If you specify a 50

Ω termination but are actually

terminating into an open circuit, the actual output will be
twice the value specified.
For example, if you set the offset to 100 mVDC (and
specify a 50

Ω load) but are terminating the output into

an open circuit, the actual offset will be 200 mVDC.

• If you change the output termination setting, the output

amplitude and offset levels are automatically adjusted (no
error will be generated).
For example, if you set the amplitude to 5 Vpp and then
change the output termination from 50

Ω to “high

impedance”, the amplitude value will double to 10 Vpp. If
you change

from “high impedance” to 50

Ω, the displayed

amplitude value will drop to half.

• You cannot specify the output amplitude in dBm if the

output termination is currently set to “high impedance”.
The units are automatically converted to Vpp.