Pulse Waveform
As shown below, a Pulse waveform consists of a period, a
pulse width, a rising edge, and a falling edge.
Figure 2-12 Pulse waveform
Pulse Period
The range of the pulse period is from 200 ns to 2000 s. The
default value is 1 ms. The U2761A will adjusts the pulse
width as necessary to accommodate the specified period.
Soft Front Panel Operation
After selecting
Pulse
at the top panel as shown in
, on the Frequency panel as shown in
select
Period
, input the desired pulse period value,
and then select the unit from the drop down list.
Remote Interface Operation
PULSe:PERiod <seconds>
Pulse width
Period
90%
90%
50%
50%
10%
10%
Rise time
Fall time