Caution, Warning – Brewer Access High-Low User Manual

Page 17

Advertising
background image

Document # 100452 RevL

17

Printed in USA © 2013

Preventative Maintenance

CAUTION

Failure to perform periodic inspections of the

table could result in personal injury or equipment

damage.

• Periodically inspect the electrical cord to ensure it is

free of cuts or damage.

• Inspect the mechanical functions to ensure satisfac-

tory operation. Complete the operating test.

• Check fasteners to make sure they are present and

tightened securely.

• Lubricate moving parts (such as back and storage

cover hinges) with a food grade silicon lubricant.

• Have a qualified technician inspect your table every 6

months.

Control Box Fuse Replacement

WARNING

Disconnect power by removing plug from the

power outlet before changing fuses.

1. Manually raise table and prop in raised position.

2. Remove shroud below seat to expose actuators and

control box. (See Figure 32.)

3. Locate fuse cover on control box. (See Figure 33.)

4. Pull out fuse cover with pliers. Note: pull cover

straight out. Do not twist.

5. Replace fuses using spare fuses in fuse cover. Fuse

type is 250V/2.5AT for 115 VAC and 250V/1.25AT for

230 VAC models.

6. Reassemble.

*Drawer Warmer Fuse Replacement

(Models 6001, 6501 and 6801)

1. Remove two fuse covers from back of table near the

power cord.

2. Replace both fuses. Fuse type is 5 x 20mm,

250V/1.0A.

3. Replace fuse covers.

* Models 6001, 6501 and 6801 with Warmer & Outlet

have a single fuse. Fuse type is 5 x 20mm, 250V/3.0A.

MAINTENANCE

CAUTION

The upholstery material is resistant to most medici-

nal type stains, but may be damaged by solvents

and dyes. Remove any spilled fluids from the up-

holstery immediately.

Clean the table weekly, wiping the painted metal and

plastic surfaces with a clean, soft cloth. If desired, mild

cleaners may be used.

Clean the work surface and rear paper storage cover im-

mediately after contact with any fluids or gels. Use a mild

soap and rinse with clear water. Avoid the use of harsh

chemicals as they can permanently damage the finish.

Table Care

Upholstery Removal

1. Remove the two screws under the seat securing the

pelvic tilt lever to the seat frame (models equipped with

pelvic tilt only, all other models skip to step 2).

2. Lift the seat and remove the three screws securing the

hinge to the seat frame.

3. Remove one screw in the paper holder securing the

seat back to the back frame.

4. Remove the upholstery from the table.

5. To remove the leg extension upholstery, pull up on pad

to release the plastic clips.

Remove Two Screws

Remove Cover

Remove Fuse Cover

and Replace Fuses

Figure 32.

Figure 33.

Advertising