Printed in USA © 2013

Overview

Perform the following sequence in order when setting up

the table:
- Uncrating

-

Leveling the Base

-

Installing Grab Bars

Uncrating

NOTE: Inspect all boxes and contents for damage. Re-

port any damage to the carrier immediately.

CAUTION

To avoid damaging the table’s upholstery or paint-

ed surfaces, DO NOT use a knife or other sharp

object to open the packaging. Also, to avoid dam-

aging the table lift only at points indicated in Figure

3. Do not lift at other points as indicated by the

note in Figure 4.

CAUTION

Do not plug power into table prior to removal of all

packaging materials.

1. Remove stretch wrap surrounding table.

2. Cut and remove banding and protective foam tube

attaching table to pallet.

3. Remove the 6 steel shipping brackets securing the table

to the pallet, one at each of the four corners and one

at the middle on each side of the table (Figure 3). The

brackets engage with the table levelers to hold the table

securely to the pallet. Remove the 2 bolts from each

bracket using a 7/16” socket or wrench. Take care not

to scratch the unit when turning the wrench. Remove

brackets from levelers once bolts are removed.

4. Remove the 2 screws attaching the spacers on either

side of the wooden U-shaped brace located under the

front corners of the cabinet (Figure 4). Remove the

2 screws attaching the U-shaped brace to the pallet

(Figure 4).

5. Remove the foot control from the rear drawer. Insert the

foot control cord into the receptacle at the bottom of the

rear panel (

Figure 10)

. Secure with locking ring.

INSTALLATION

WARNING

The table weighs approximately 395 lbs. Use

enough personnel to remove the table from the

shipping pallet. Also, use proper lifting techniques

when lifting. Failure to do so could result in seri-

ous injury.

CAUTION

Do not use the stirrups to lift or move the table.

The Leg Support may be used to lift the table. Do

not pull on the Leg Support to slide the table.

See

SPECIFICATIONS

table for electrical rating of

this unit. The three-pronged grounding plug on the table

power cord must be plugged into a matching three-

pronged, grounded, non-isolated, correctly polarized 115

VAC receptacle.

CAUTION

Do not lift table by seat upholstery. Damage to

the pelvic tilt mechanism and/or upholstery may

occur.

Figure 3. Remove Shipping Brackets

Figure 4. Remove Spacers and U-Brace

Advertising