3.1.4 Battery Charging

To check the level of charge of the battery, press the button on the battery’s keypad. The battery
gauge/indicator(s) illuminate to indicate the level of battery charge (25-100%).

3.1.4.1 To charge the Focus battery while using the power supply with Focus

1a) Using AC Power: Follow the instructions in the Connecting Focus to AC electrical power
section.

1b) Using DC Power: Follow the instructions in the Connecting Focus with the Battery section.

2) Connect the battery cord without the on / off switch into the power supply outlet marked DC
OUT To Battery. Connect the other end to the battery.

Note: The battery is charging whenever the unit operates on AC or DC power.


The Focus battery will completely recharge from its fully depleted state in

approximately four hours, whether or not the unit is in use on AC or DC power.

While charging a fully discharged battery, the LED will continue to blink until 25%

capacity is reached. The LED will then turn solid.

Each of the four LEDs, 25% -100%, will blink as stated above, then turn solid when

they reach their capacity.

When all LEDs illuminate solid, the battery is fully charged and the LEDs will remain

solid for a period of time, then all four LEDs will turn off

.

Wherever storing Focus battery for an extended period of time, ensure that the battery state of

charge (SOC) is not depleted or stored at 100%. Both these conditions can result in
unrecoverable capacity loss. The recommended SOC is 25% to 50% for long-term storage.
The recommended storage temperature is 23

o

C +/- 2

o

C.


















The Battery can be recharged from the power
supply

while using Focus AC or DC power.