3.1.4 Battery Charging
To check the level of charge of the battery, press the button on the battery’s keypad. The battery 
gauge/indicator(s) illuminate to indicate the level of battery charge (25-100%). 
3.1.4.1 To charge the Focus battery while using the power supply with Focus
1a) Using AC Power: Follow the instructions in the Connecting Focus to AC electrical power 
 section. 
1b) Using DC Power: Follow the instructions in the Connecting Focus with the Battery section. 
 
 2) Connect the battery cord without the on / off switch into the power supply outlet marked DC 
 OUT To Battery. Connect the other end to the battery. 
Note: The battery is charging whenever the unit operates on AC or DC power.
 
• The Focus battery will completely recharge from its fully depleted state in 
approximately four hours, whether or not the unit is in use on AC or DC power.
• While charging a fully discharged battery, the LED will continue to blink until 25%
capacity is reached. The LED will then turn solid.
• Each of the four LEDs, 25% -100%, will blink as stated above, then turn solid when
they reach their capacity.
• When all LEDs illuminate solid, the battery is fully charged and the LEDs will remain
solid for a period of time, then all four LEDs will turn off
.
• Wherever storing Focus battery for an extended period of time, ensure that the battery state of
charge (SOC) is not depleted or stored at 100%. Both these conditions can result in 
unrecoverable capacity loss. The recommended SOC is 25% to 50% for long-term storage. 
The recommended storage temperature is 23
o
C +/- 2
o
C.
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
The Battery can be recharged from the power 
supply
while using Focus AC or DC power.
 
 
F
i
g
u
r
e
 
4
 
Figure: 11 
F
i
g
u
r
e
 
4