AirSep MN171-1 User Manual
Page 8
 
MN171-1 Rev. A 02/14
1.4 Focus Specifications
Oxygen Concentration:*
90% oxygen - 3% / +5.5%
 
Dimensions: 
 
6.4 in. high x 4.8 in. wide x 2.5 in. deep (16.4 cm high x 12.2 cm wide x 6.1 deep) 
Weight:
Concentrator 1.75 lb (0.8 kg)
Battery 0.53 lb (0.2 kg) 
Optional AirBelt Battery 1.8 lb (0.8 kg) 
 
Power: 
Universal Power Supply: 
AC Power Supply: Input # 1_100 – 240 VAC (1.5 Amps max at 120 VAC 50/60 Hz) 
DC Power Supply: Input # 2_11-16 VDC 5.0 Amps max 
 
Battery duration 
(Rechargeable 
 lithium Battery) 
 
Battery: 1 ½ hours (per battery) 
Optional AirBelt Battery: 4 hours 
 
 
Battery recharge time: 
4 hours; optional AirBelt: 3 ½ hours
Warm-up time:
2 minutes
 
Battery cycle life: 
Approximately 300 cycles, then 80% capacity or below.
Audible alarms and pulse 
visual indicators: 
Start-up – audible and visual 
Pulse flow – visual 
Cannula disconnect- audible and visual 
Breath rate alarm – audible and visual 
General malfunction – audible and visual 
Service required - visual 
Low battery - audible and visual
Battery condition – battery level indicator on battery
Temperature range:
Normal operating temperature: 41°F to 104°F (5°C to 40°C) 
Storage temperature: 
-2°F to 140°F (-20°C to 60°C)
 * Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C) 