Problem possible cause solution – AirSep MN171-1 User Manual
Page 23
                                    Advertising
                                    
                                        
                                    
                                
                                 
MN171-1 Rev. A 02/14
Problem
Possible Cause
Solution
Battery does not charge.
Battery temperature has exceeded 
battery operating temperature 5 to 
40
o
C.
 
Battery is defective. 
 
AC adapter connector is defective or 
contact pin is bent. 
 
Defective AC adapter 
Allow battery to cool before 
charging. 
 
 
Replace battery 
 
Replace battery 
 
 
Replace AC adapter. 
Oxygen concentration is below 
specifications in test mode. 
Unknown 
 
 
Return to AirSep. 
 
 
Unit does not deliver oxygen.
Cannula kinked or not properly 
connected. 
 
Unknown 
Check cannula and the 
cannula connection. 
 
Return to AirSep. 
 
 
 
 
 
 
 
 
 
 
 
 
 