Preparing the keepsafe essential for use, Keepsafe, Essential 7 – Posey KeepSafe® Essential User Manual

Page 7: Battery installation

Advertising
background image

KeepSafe

®

Essential

7



Preparing the KeepSafe Essential for Use

Battery Installation:

The battery operated Posey KeepSafe Essential is portable and long lasting. Fresh alkaline batteries have an

estimated life of 30 days of daily use. Actual life depends on alarm mode, tone and volume you select.
Alarm will “chirp” about every 15 seconds when new batteries are

needed. Change batteries at once.
1. If your alarm does nOT have an On/OFF switch (Cat. 8373nP),

connect the magnet before inserting the batteries. If your alarm dOES

have an On/OFF switch (Cat. 8373), slide the power switch to the OFF

position before inserting/changing batteries.

2. Press down on the arrow and slide the battery compartment door

completely off (fig. 1). Set battery door aside.

3. drape the red ribbon across the battery compartment and insert four

(4) new “AAA” batteries as pictured inside the battery compartment

(fig. 2). Take care not to damage battery contacts. Batteries should be

inserted on top of red ribbon and the loose end of the ribbon should

stick out the other end for easy battery removal (fig. 3).

4. Lay the loose end of the red ribbon on top of the batteries and

reattach the battery compartment door. Slide it shut, locking it into

place.

• Take care when installing new batteries. The alarm will not work if

batteries are installed improperly.

• ALWAYS install a completely new set of batteries when the

chirping due to low battery power sound begins. DO NOT replace

a single cell, but all cells in the alarm. DO NOT mix old and new

batteries, or battery brands within a battery pack (4 batteries).

Use of mixed batteries or batteries installed incorrectly may cause

battery damage, and may damage the alarm. Remove any alarm

from use and send to the appropriate facility authority if batteries

are damaged or corroded or the battery compartment has signs of

previous battery corrosion such as white powder residue.

• Batteries can explode or leak and cause damage to alarm if installed incorrectly, fully discharged,

or exposed to liquid, fire or high temperatures. If battery damage has occurred, or you see any

corrosion, remove the alarm from use IMMEDIaTELy. DO NOT use the alarm if battery damage has

been detected.

• After changing batteries, test the alarm for proper operation prior to putting in service with a

patient, and each time before leaving the patient unattended. If the alarm does not function

properly, remove the alarm from service and replace it with a properly functioning alarm (see

pages 17-18).

Fig. 2

Fig. 1

Fig. 3

Advertising