Keepsafe, Deluxe 15 – Posey KeepSafe® Deluxe User Manual

Page 15

Advertising
background image

KeepSafe

®

Deluxe

15



Nurse Call Interface (with 8374NC or 8374NPNC System)

You can connect the KeepSafe Deluxe Nurse Call System to your facility nurse call system. This allows you

to add the options and notification associated with your specific nurse call system such as nursing station

notification and patient room lights and sounds.

To Connect the Nurse Call Cable:

1. Insert one end of cable into “Nurse Call” jack on left side of alarm (fig. 10).

2. Insert the other end of cable into wall jack of nurse call panel.

Depending on your system, you may need a “Y” adaptor to use the

Posey KeepSafe Deluxe Nurse Call System and the patient nurse call

device at the same time.

When using a nurse call cable, ensure the nurse call

cable is plugged in to both the alarm and the wall jack before leaving

the patient unattended. Verify that an alert is received at the nursing

station if the cable is unplugged from the wall jack.
NOTE: There will be no alert at the nursing station or at the bedside if the nurse call cable is

unplugged from the alarm.

FOR SaFE uSE WiTh NuRSE Call CaBlE:

• DO NOT stretch or strain cable to avoid possible damage and possible malfunction.

• DO NOT attach cable to moving parts of the bed or chair that will cause strain or damage if the

bed or chair is repositioned.

• ALWAYS position the cable so that moving parts (side rails, wheels, etc.) will not cause strain or

damage the cable.

• DO NOT run over the cable with carts or equipment.

• DO NOT wrap the cable tightly during storage.

• ALWAYS remove the cable by pulling on the jack. DO NOT pull on the cable.

• alWayS secure the cable out of the way so it will not be a tripping hazard.

• ALWAYS test alarm and nurse call function prior to leaving the patient unattended. Activate the

alarm (remove pressure from sensor, unfasten chair belt sensor, or activate PiR sensor) and make

sure the nurse call light for the proper bed and room activate in the hall and at the nurse’s station.

• DO NOT use the alarm or sensor if it does not activate each time weight is removed from the

sensor, the chair belt sensor is unfastened, or the PiR sensor is activated. Replace alarm and sensor

with working units and retest before leaving the patient unattended (see pages 19-20).

9VDC

Fig. 10

Advertising