Setting alarm mode, Keepsafe, Deluxe 9 – Posey KeepSafe® Deluxe User Manual

Page 9: The pre-recorded voice message is, Mode features, Tone

Advertising
background image

KeepSafe

®

Deluxe

9



When using a nurse call cable, ensure the nurse call cable is plugged in to both the alarm

and the wall jack before leaving the patient unattended. Verify that an alert is received at the nursing

station if the cable is unplugged from the wall jack.
NOTE: There will be no alert at the nursing station or at the bedside if the nurse call cable is

unplugged from the alarm.

To Change/Select Mode:

Follow these steps to change or select alarm mode:
1. Ensure alarm is on and either sensor pad, chair belt sensor or PIR

sensor is attached to alarm. Ensure pressure is on sensor pad, chair

belt sensor is connected, or PIR sensor is connected.

2. Remove the battery compartment door by sliding down to gain access

to the programming buttons.

3. Slide the MoDE Select Switch to the desired alarm mode (fig. 4).
The last option selected is the mode utilized when the sensor is

activated.
4. Replace the battery door and slide shut, locking into place.
NOTE: Cat. 8374NP does NoT have an oN/oFF switch. To set Alarm Mode for Cat. 8374NP, follow all steps

above except step 1.

The Pre-Recorded Voice Message Is:

“Please don’t get up. Sit back down and use the call button to call for help.”

Setting Alarm Mode

The KeepSafe Deluxe has three (3) alarm modes. These allow you to select an alarm signal best suited to

patient and facility needs.

Mode Features:

Fig. 4

Tone

Volume

Voice

&

Tone

Voice

Tone

Qt

y

4A

AA

A

lk

ali

ne

AN

SI-

24

A I

EC

-L

R0

3

VOiCE & TONE:

Pre-recorded voice message plays once, followed by the selected alarm tone. The

tone continues to play until alarm is placed on hold or patient is repositioned with

the sensor.

TONE:

Selected alarm tone plays until alarm is placed on hold or patient is repositioned

with the sensor.

VOiCE ONly:

Pre-recorded voice message plays until alarm is placed on hold or patient is

repositioned with the sensor.

Advertising