Calibration, Indication of the battery voltage – Vaisala Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 User Manual

Page 11

Advertising
background image

HMI41 Calibration Option
Appendix 1

19165ZZ-U220en-1.1

2

1996-08-26

2.

Calibration

After having connected the cable, turn the HMI41 on and follow the
instructions of the table below:

DISPLAY

WHAT TO DO

PRESS:

RH

setminmax hyst hold Lo bat Hi

Pabs

Pa

g/m

T

Td

°C°F

°C°F

%

The HMI41 has been turned on.

Indication of the software version (if the
version is 1.02 or more)

bat Hi

Indication of the battery voltage.

RH

%

Let the readings stabilize and then adjust
the transmitter potentiometers until the
readings are the same.
offset: <65 %RH
gain : >65 %RH
Alternatively, you can adjust the
transmitter when the display shows the
difference in readings.

MODE to have the difference
in readings on the display
ON/OFF if you have
completed the calibration

RH

%

Let the readings stabilize and then adjust
the transmitter potentiometers until the
reading on the first line is as close to zero
as possible.
offset: <65 %RH
gain : >65 %RH

MODE to return the measured
values on the display
ON/OFF if you have
completed the calibration