HMI41 Calibration Option

19165ZZ-U220en-1.1

Operating Manual

1996-08-29

3

2.

CALIBRATION

For a successful calibration, it is essential that the probe of the HMI41 and
that of the transmitter are at the same temperature, and that the reference
probe has been previously calibrated. Always allow enough time for the
readings to stabilize. Note that the stabilization time depends on the ambient
conditions and may vary from 10 minutes to a couple of hours.

2.1

Getting started

After having selected the desired function (Chapter 1.2), turn the HMI41 off
and connect the calibration cable to the EXT connector at the bottom of the
HMI41 (Figure 2.1). Connect the other end of the cable to the appropriate
connector in the transmitter (Figures 2.2 - 2.4).

EN TER

MO DE

HOLD

ON/OFF

EXT

C o n n e c t o r f o r t h e

c a l i b r a t i o n c a b l e

Figure 2.1

Location of the HMI41 calibration connector

one point
RH calibration

test connector

Figure 2.2

Calibration connector and potentiometer in

the HMP130 series transmitters