2 calibration – Vaisala Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 User Manual

Page 8

Advertising
background image

HMI41 Calibration Option
Operating Manual

19165ZZ-U220en-1.1

4

1996-08-29

offset

gain

test
connector

Figure 2.3

Calibration connector and potentiometers in

the HMD 20/30UB/YB transmitters

gain

offset

test
connector

Figure 2.4

Calibration connector and potentiometers in

the HMW 20/30UB/YB transmitters

2.2

Calibration

After having connected the cable, turn the HMI41 on with the ON/OFF button
and wait until a text similar to the following appears on the display:

RH

%

Numbers on the first line indicate the RH measured by the transmitter, and
numbers on the second line indicate the RH measured by the HMI41 reference
probe. If you wish, you can now adjust the transmitter:

Wait until the readings have stabilized and adjust the calibration

potentiometers (Figures 2.2 - 2.4) until the readings on the first and
the second line are the same. It is recommended to use the offset
potentiometer if the RH is <65 %RH and the gain potentiometer if the
RH is >65 %RH.