Vaisala Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 User Manual

Page 6

Advertising
background image

HMI41 Calibration Option
Operating Manual

19165ZZ-U220en-1.1

2

1996-08-29

set

°C

If the basic settings of the HMI41 (display units, automatic power off
function, display quantities and pressure) have to be changed, please refer to
the HMI41 Operating Manual. Otherwise, press ENTER repeatedly until the
following text appears:

set

The number on the first line of the display corresponds to the following
HMI41 functions:

1 = HMI41 as a humidity and temperature indicator
2 = HMI41 as a calibrator for HMD/W 60/70 and HMP140 series analogue
transmitters
3 = HMI41 as a calibrator for HMP230 series digital transmitters
4 = HMI41 as a calibrator for the HMD/W20/30 and HMP130 series
analogue transmitters

To calibrate the HMD/W 20/30 or HMP130 series transmitters, select number
4 with buttons ▲ (number up) and ▼ (number down) and then turn the HMI41
off. When the HMI41 is turned on again, it will automatically assume the
calibrator function for the HMD/W 20/30 and HMP130 series transmitters. If
you wish to calibrate transmitters of some other type or use the HMI41 as an
indicator, this selection has to be made accordingly; refer to corresponding
manuals for further details.