2 temperature channel – Vaisala Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 User Manual
Page 13

HMI41 Calibration Option
19116ZZ-U218en-1.1
Operating Manual
1996-08-29
9
2.3.3.2
Temperature channel
T
set
Numbers on the second line are blinking. They indicate the voltage range of
the transmitter’s T channel (jumper selectable) currently stored in the HMI41
memory. Please, consult Figures 2.3.3 - 2.3.4 when checking the transmitter’s
jumper positions.
T output
selections
0...1V
0...5 V
0...10 V
Figure 2.3.3
Jumper selections for the HMD/W70
T output
selections
0...1V
0...5 V
0...10 V
Figure 2.3.4
Jumper selections for the HMP140
You can now set the range with buttons ▲ or ▼ to correspond to the jumper
position. Acknowledge the setting with ENTER. A text similar to the
following is displayed:
set
Lo
T
Numbers on the second line are blinking. They indicate the lower limit of the
transmitter’s voltage range currently stored in the HMI41 memory. Text in the
lower righthand corner is Lo. You can now set the lower limit with buttons ▲
or ▼ to correspond to that of the transmitter. Acknowledge the selection with
ENTER. A text similar to the following is displayed: