1 humidity channel – Vaisala Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 User Manual

Page 15

Advertising
background image

HMI41 Calibration Option

19116ZZ-U218en-1.1

Operating Manual

1996-08-29

11

Select the cI selection mode with buttons ▲ or ▼ and acknowledge the
selection with ENTER. A text similar to the following is displayed:

set

RH

2.3.4.1

Humidity channel

set

RH

Numbers on the second line are blinking. They indicate the current range of
the the transmitter’s RH channel currently stored in the HMI41 memory.
Select the range with buttons ▲ or ▼ to correspond to that of the transmitter.
Acknowledge the selection with ENTER. A text similar to the following is
displayed:

RH

set

Lo

%

Numbers on the second line are blinking. They indicate the lower limit of the
transmitter’s RH range currently stored in the HMI41 memory. Text in the
lower righthand corner is Lo. You can now set the lower limit to correspond to
that of the transmitter. Set the limit with buttons ▲ or ▼ and acknowledge the
setting with ENTER:

RH

set

Hi

%

Numbers on the first line start to blink. They indicate the upper limit of the
transmitter’s RH range currently stored in the HMI41 memory. Text in the