HMI41 Calibration Option

19116ZZ-U218-1.1

Appendix 1

1996-08-26

5

2.4

Customized current output selection mode (cI)

DISPLAY

WHAT TO DO

PRESS:

Select the customized current output scale
selection mode (cI)

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection

set

RH

Select the RH channel’s current output
range
(4...20 mA, 0...20 mA)

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection

RH

set

Lo

%

Set the lower limit of the humidity range

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection

RH

set

Hi

%

Set the upper limit of the humidity range

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection

set

T

Select the T channel’s current output
range
(4...20 mA, 0...20 mA)

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection

set

Lo

T

T

°C

°C

Set the lower limit of the temperature
range

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection

set

Hi

T

T

°C

°C

Set the upper limit of the temperature
range

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
press ON/OFF