4 customized current output selection mode (ci) – Vaisala Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 User Manual
Page 26

HMI41 Calibration Option
19116ZZ-U218-1.1
Appendix 1
1996-08-26
5
2.4
Customized current output selection mode (cI)
DISPLAY
WHAT TO DO
PRESS:
Select the customized current output scale
selection mode (cI)
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
RH
Select the RH channel’s current output
range
(4...20 mA, 0...20 mA)
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
RH
set
Lo
%
Set the lower limit of the humidity range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
RH
set
Hi
%
Set the upper limit of the humidity range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
T
Select the T channel’s current output
range
(4...20 mA, 0...20 mA)
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
Lo
T
T
°C
°C
Set the lower limit of the temperature
range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
set
Hi
T
T
°C
°C
Set the upper limit of the temperature
range
▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection
press ON/OFF