HMI41 Calibration Option
Appendix 1

19116ZZ-U218en-1.1

2

1996-08-26

2.

Scale selection modes

The HMI41 automatically assumes the function selected in the setup menu
when it is turned on again with the ON/OFF button. In the following, you will
find a summary of the scale selection modes.

DISPLAY

DESCRIPTION

RH

setminmax

hyst hold Lo batHi

Pabs

Pa

g/m

T

Td

°C°F

°C°F

%

The HMI41 has been turned on.

Indication of the software version (if the version
is 1.02 or more)

bat Hi

Indication of the battery voltage.

RH

T

Press any button except ON/OFF within 5
seconds to enter the scale selection modes.

Select the desired scale selection mode with
buttons ▲or ▼ and acknowledge the selection
with ENTER. Then follow the instructions of
tables 2.1 - 2.5 accordingly.

2.1

Voltage output scale selection mode (U)

DISPLAY

WHAT TO DO

PRESS:

Select the voltage output scale selection
mode (U)

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection

set

Lo

T

T

°C

°C

Set the lower limit of the temperature
range

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection

set

Hi

T

T

°C

°C

Set the upper limit of the temperature
range

▲ (number up) or
▼ (number down)
ENTER to acknowledge the
selection