HMI41 AND HMP42

M210317EN-A

Operating Manual

11

5.

CALIBRATION

5.1

How to use the HMP42 calibration adapter

As the HMP42 probe diameter is extremely small, the probe has a special
calibration adapter for inserting the probe into a salt chamber (e.g. in the
HMK15 calibrator).

NOTE

Be extremely careful when inserting the probe into the
calibration adapter: do not bend or break it.

The probe is inserted into the adapter with the steel grid on; this protects the
sensor and prevents it from breaking. Unlike other humidity probes, the
HMP42 is calibrated with the steel grid on and thus, the probe naturally
requires a slightly longer stabilization time (approx. 30 minutes).

Figure 5.1.1

HMI41, HMP42 and calibration accessories

Figure 5.1.2

Insert the probe head into the calibration adapter