Vaisala HMP42 User Manual

Page 19

Advertising
background image

HMI41 AND HMP42

M210317EN-A

Operating Manual

15

The calibration procedure is included in the HMI41 setup mode. For entering
the setup mode, first press the ON/OFF button and the following appears:

RH

set min max

hyst hold

Lo bat Hi

P

Pa

h

g/m

T

Td

°C °F

°C °F

%

Then release the ON/OFF button and press within 1...2 seconds both ENTER
and MODE buttons until the following text appears on the display:

After a few seconds, the text changes automatically to show the following:

set

°C

Press ENTER eight times until the following text appears:

set

This indicates that no calibration has been selected. In addition to this, there
are five calibration types available. The desired type is selected with buttons
▲ or ▼.

Advertising
This manual is related to the following products: