4 two-point temperature calibration, Press enter eight times, Press mode five times – Vaisala HMP42 User Manual

Page 62

Advertising
background image

HMI41 AND HMP42
Operating Manual

M210317EN-A

58

2.4

Two-point temperature calibration

RH

set min max

hyst hold

Lo bat Hi

P

Pa

h

g/m

T

Td

°C °F

°C °F

%

Turn the HMI41 indicator on. Within 1-2 seconds
press simultaneously buttons MODE and HOLD
until the text SEtUP appears.

Wait a few seconds.

set

°C

Press ENTER eight times.

set

Press MODE five times.

set

T

Press ENTER to activate two-point temperature
calibration.

set

Lo

T

°C

The lower reference reading stored in the HMI41
memory is blinking. Change it to the correct value
with buttons ▲ or ▼. Press ENTER.

Lo

T

°C

The indicator shows the value that the probe is
currently measuring. Wait at least 10 minutes for
the reading to stabilize and acknowledge it with
ENTER. Press ENTER again to conclude the lower
point calibration.

set

Hi

T

°C

The higher reference reading stored in the HMI41
memory is blinking. Change it to the correct value
with buttons ▲ or ▼. Press ENTER.

Hi

T

°C

The indicator shows the value that the probe is
currently measuring. Wait at least 10 minutes for
the reading to stabilize and acknowledge it with
ENTER. Press ENTER again to conclude the
calibration.
Calibration was successful. If it was not succesful,
some other text appears on the display (e.g. too
close, err offst
or err gain). Perform the calibration
again.