Set analog output scaling – Vaisala HMW90 User Manual

Page 50

Advertising
background image

User's Guide _______________________________________________________________________

48 ___________________________________________________________________ M211399EN-F

Set Analog Output Scaling

Use the ASEL command to select the output parameter and scaling for

analog output channels.

ASEL [channel parameter lo_value hi_value]<cr>

where
channel = Analog output channel, 1 or 2.

parameter = Parameter that is output on the channel.

Available parameters are:

RH

relative humidity

T

temperature

Tdf

dew/frost point temperature

Td

dew point temperature

Tw

wetbulb temperature

h

enthalpy

x

mixing ratio

a

absolute humidity

dTd

dew/frost point depression

lo_value = Low limit of the scaling, in the units of the selected

parameter.

hi_value = High limit of the scaling in the units of the selected

parameter.

Example (set channel 1 to output dewpoint, in the range -10 ... 20 °C):

>pass 9000

>asel 1 TD -10 20

Aout 1 quantity : Td (-10.00 ... 20.00 'C)

Advertising